Cookie Settings

We use cookies to optimize our website. These include cookies that are necessary for the operation of the site, as well as those that are only used for anonymous statistic. You can decide for yourself which categories you want to allow. Further information can be found in our data privacy protection .

Essential

These cookies are necessary to run the core functionalities of this website and cannot be disabled.

Name Webedition CMS
Purpose This cookie is required by the CMS (Content Management System) Webedition for the system to function correctly. Typically, this cookie is deleted when the browser is closed.
Name econda
Purpose Session cookie emos_jcsid for the web analysis software econda. This runs in the “anonymized measurement” mode. There is no personal reference. As soon as the user leaves the site, tracking is ended and all data in the browser are automatically deleted.
Statistics

These cookies help us understand how visitors interact with our website by collecting and analyzing information anonymously. Depending on the tool, one or more cookies are set by the provider.

Name econda
Purpose Statistics
External media

Content from external media platforms is blocked by default. If cookies from external media are accepted, access to this content no longer requires manual consent.

Name YouTube
Purpose Show YouTube content
Name Twitter
Purpose activate Twitter Feeds

RF Arrays

Radiofrequency (RF) arrays play a crucial role in MRI. In all modern MRI systems, arrays of RF coils are used to detect the signals emanating from the spins in the human body. Larger number of elements in such receive arrays can increase the SNR and allow for higher acceleration of the acquisition.

While clinical MRI systems at 1.5T and 3T use birdcage volume coils for transmission and arrays for reception, at ultra-high field (UHF) MRI, RF arrays are not only used for reception, but also for transmission. This is to gain more degrees of freedom when shaping the transmit field, as volume coils do not produce a homogenous excitation field at UHF. Similar to beam forming in modern radar technology, the elements of a transmit array in MRI are driven with individual amplitudes and phases and even different pulse shapes.

These arrays can be characterized be a multitude of criteria. They should have a high transmit efficiency to provide a strong excitation field with the transmit power provided by the system, as well as a high SAR efficiency to allow for a strong excitation field without damaging tissue due to excessive heating. Furthermore the elements of the array need to be well decoupled and their respective transmit profiles need to be distinct from one another to provide a high degree of freedom for shaping the transmit field.

Left: 32-channel transmit array for large field of view imaging used at the DKFZ's 7T MRI. The array is mounted on the bore liner. Right: Inside view of one half of the array. 16 transmit elements in 3 rows are visible.
© dkfz.de

While transmit RF arrays at UHF are almost exclusively local arrays placed directly on the body, our group aims at developing RF arrays that are integrated in the MRI system behind the inner cladding of the bore, just like the transmit volume coils of lower field MRIs [1].


Our current and future work in this area includes:

  • Finding transmit antennas best suited for integrated arrays
  • Developing very thin arrays to fit between boreliner and gradient coil of the most modern MRI systems
  • Developing meta material solutions for improved coil performance

 